7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 41 of 71

Calibrating the Temperature

If temperature calibration is needed, the following steps can be taken:

Locate the reference instrument’s temperature sensor in close proximity to the

cabinet’s geometric center. Be sure that the stabilization times described earlier have

been satisfied prior to performing calibration.

At the calibrate screen, press the

(Temperature Calibrate) button.

Enter the temperature offset by using the keypad and pressing

(Enter) when

complete.

A positive value will move the temperature ‘up’ and a negative value ‘down’. Press the

‘home’ button and verify the proper temperature is displayed.