7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 44 of 71

Snoozing the speaker:

When in an alarm condition, the speaker can be temporarily silenced to avoid being a

nuisance to those nearby. The alarm will reappear after 60 minutes has passed, if the

condition has not been corrected. (The audible alarm will not sound if the alarm

speaker is muted)

Press the

(Snooze) button, the audible alarm is silenced for a period of 60

minutes.

When the alarm condition is corrected the alarm indicator and the audible alarm will

automatically turn off (unless there is another alarm condition).

To check what the alarm condition is, press the

(Alarm) button on the status bar.

and the alarm window will be displayed. If the

(Snooze) button has already

been pushed and 60 minutes have not passed the Snooze button will be “greyed” out.

If you press

the

(Close Window) button, the Alarm Window will close, but the

alarm will still be present as a flashing alarm icon on the status bar for the remainder of

the 60 minutes time. It will not reset the 60 minutes alarm countdown time if the alarm

condition is viewed on the pop up window.