Caron 7000-75 User Manual

Page 13

Advertising
background image

7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 13 of 71

Connect each chamber to a grounded circuit.

Failure to do so could result in electrical shock.

Use only distilled or deionized water with a resistivity

between 50K

-CM and 1M

-CM and a pH of greater than

6.5. Using water outside this range will void your warranty.

Do not use water that contains chloramines. Chloramines

can damage internal rubber gaskets resulting in leaks.

Connecting the Water Supply

To ensure proper operation, distilled or deionized water is required as a supply on units

that have humidity control. If these water sources are not available contact CARON

customer service.

A water inlet fitting on the back of the unit and ¼” black tubing are provided to connect

the water supply to the chamber. Connect an appropriate water supply to the fitting.

Incoming line pressure should be regulated to not exceed 80 psi.

If a Condensate Recirculator water recycling system was purchased as a water supply,

refer to its user’s manual for proper installation of the water supply.

Connecting Electrical Power

The unit requires a dedicated electrical outlet. See table below for model specific

power required and connection.

Model #

Power Requirements

Plug Connection

-1

115V, 60Hz, 16A FLA

NEMA 5-20

-2

230V, 60Hz, 12A FLA

NEMA 6-15

-3

230V, 50Hz, 10A FLA

CEE 7/7

When the required electrical connection is available, plug the provided power cord into

the unit and the electrical outlet.

Advertising
This manual is related to the following products: