Despatch Protocol Manager with Protocol Plus User Manual

Page 27

Advertising
background image

Protocol Manager Software E-100

23

Double-click on the fields in the center of the Recipe/Datalog window to bring up a
window box to access and change the value shown. Move the slider to change the
value quickly. Click on the arrow at either end to change the value by 1. Click in the
white space to the left and right of the slider to change the value by 10. Click OK to lock
in the change. See Figure 22.

Figure 22: Value Change Window


Parameters on the Recipe screen include:

Recipe Name

Type the name of the new recipe being created.

Seg#

Recipe segments 1 through 8 may be programmed, each with its
own set of events, ramp and soak times, and soak temperature.

Ramp Time

The time required to ramp setpoint up to temperature may be
programmed here. Values between 0 and 99:59 are allowable. As in
the Protocol Plus controller, profile ramp and soak times are stored
without units. Units are independently set locally, in the actual
controller, as either hours and minutes (HH:MM) or minutes and
seconds (MM:SS).

EV1 through 4 From 1 to 4 events may be programmed into the ramp time portion

of each segment here. These typically involve actuating/disabling
relays to close/open valves or perform other relay-controlled
functions. NOTE: These will only actuate when the controller has the
relay cards installed and programmed for an event. Recipes may
contain four Ramp and four Soak event choices, although only those
relays actually supplied in a Protocol Plus controller and configured
for events will be affected by recipe settings, when used to run a
profile.

Soak Temp.

The temperature of a particular segment is entered here; it can
range from -18 to 540 degrees C (0 to 1000 degrees F). See Notes
on Soak Temp and Hi Limit SP, below.

Soak Time

The duration of soak is entered here; the value can range from 0 to
99:59.