Troubleshooting the communications option, Controller parameters, Protocol manager parameters – Despatch Protocol Manager with Protocol Plus User Manual

Page 52

Advertising
background image

Protocol Manager Software E-100

48

TROUBLESHOOTING THE
COMMUNICATIONS OPTION

Controller Parameters

(see Protocol Plus manual to enter Communications Page)

Address (Default = 1) This setting must be different on each control/oven if using

RS422/RS485 interface.

Mode (Default = OFF) This must be set to Modbus.
Baud Rate (Default = 19.2K) If you are having problems going to a lower baud

rate may help, but do not change to start with.

Parity (Default = None) This should remain the default.


Protocol Manager Parameters

1. Log on the highest level security (Level 4). The default password initially is

“despatch”.

2. Click on the Windows pulldown menu at the top of the screen.

3. Click on Setup menu.

4. Click on the Comm Setup tab and verify the settings match the controller

parameters set above. Click on the OK button when finished.