Protocol Manager Software E-100

24

EV1 through 4 From 1 to 4 events may be programmed into the soak portion of

each segment here. These typically involve actuating/disabling
relays to close/open valves or perform other relay-controlled
functions. NOTE: These will only actuate when the controller has the
relay cards installed and programmed for an event.

Guaranteed
Soak Band

If the process temperature deviates from the setpoint by more than
this value, the timer is placed in a hold condition. The timer
continues when the process temperature falls within range.

Hi Limit SP

The high limit setpoint may be entered here; if the temperature
exceeds this value, the hi-limit will alarm and shut off the heater. See
Notes on Soak Temp and Hi Limit SP, below.

Loop from
Segment ___
to Segment
___ ___ times

These values enable the profile or controller to jump from a certain
step to another step of the recipe a preset number of times.

Link to Profile

Values for the end of profile option are:
STOP Stop oven operation
HOLD Hold setpoint
1

– 8 Jump to selected profile number

STBY Enter standby mode. See note on Standby Mode below.

Temp. Units

Selects degrees Fahrenheit or Centigrade, for the recipe display
only. Recipes stored on the host PC, like profiles in the Protocol
Plus, are always stored with temperatures in degrees Fahrenheit,
regardless of the selected viewing units. Note that changing the
display units converts displayed recipe temperatures accordingly.

OPEN/SAVE
(to Folder)

These buttons enable the operator to open a new recipe, or save the
current one to a predetermined folder.

READ/WRITE
(to Controller)

These buttons enable the operator to upload a recipe from one of
the Protocol Plus controllers in the system, or to write the current
recipe to one of the controllers.

Profile
Number

Selects which of the Controller's eight profiles is used when reading
or writing recipes.


Notes on Soak Temp. and Hi Limit SP Values


In the recipe editor screen, Soak Temp and Hi Limit SP values can be adjusted in the
range of -18 to 538 C (0 - 1000 F). Errors may occur attempting to download a recipe
to a Controller's profile if one of the recipe's ramp times, soak temperatures, or the high
limit setpoint value is incompatible with the range set internally for the target controller.
In this case, an error message similar to Figure 23 will be displayed, and no changes
will be made to the Controller's profile.