11

MaIntenance

MediMate cga 870 and cga 950

The connector must remain clean to allow for proper operation.

Figure 25. Press on the piston. If it does not spring back, the internal components may be clogged

with soap residue. Approximately 33lbs will be required to depress the piston.

Piston

Figure 26. Remove cartridge assembly and clean residue

with water and agitation.

Figure 27. Example of contamination.

Figure 28. Check condition of clevis pin and retaining ring.

If clevis pin is broken or retaining ring is not
functional, contact factory.

Retaining Ring

Clevis Pin

Cartridge Assembly