FasTest, Inc. 1646 Terrace Drive Roseville, MN 55113 Support: (800) 444-2373 Fax: (651) 645-7390 www.fastestinc.com

6

Medical Compressed Gas Connectors

MedIcal and IndustrIal cga 540 operatIon

Safety features:

• The CGA 540 Medical and Industrial connectors have an internal locking

mechanism. Once the pressure exceeds 60 psig, the connector will lock the sleeve
to prevent removal.

Figure 13. Medical and Industrial CGA 540.

step 1

At the start of each shift:

• Check all connectors for main seal condition.
• Check for smooth operation of the sleeve and collets before the first fill.

Main Seal

Sleeve

Collet

g540-041-sH

g540-042-sH

1/4” NPT Male Incoming
Line Termination

Main Seal

Sleeve

Collet

1/4” NPT Female Incoming
Line Termination

Advertising