Medical Compressed Gas Connectors

InstallatIon

step 1

Protect the connector from damage and dirt by keeping it in the original packaging until you are ready to use it.

step 2

Check that the connector part number and delivery notes (if applicable) comply with the application.

step 3

Connect the hose securely and leak-tight to inlet or
outlet. Tighten to a max torque of 15 ft-lbs.
A higher torque can result in damage causing leaks when
the system is pressurized. Ensure that the connectors
cannot be damaged when loading and removing the
cylinder (Figure 1).

step 4

Review total connector function:
MediMate cga 870 and cga 950 (Figure 2)

• Check leak-tight seal.
• Check handle operation.
• Check seal-face/piston movement.
• Check that marking complies with application.

Medical and Industrial cga 540 (Figure 3)

• Check leak-tight seal.
• Check that collets open and close properly by

actuating the connector several times.

• Check that marking complies with application.

note
Avoid lateral forces like short connecting hoses
as this could cause leakage.

Figure 1. Torque to maximum 15 ft-lb.

Figure 2.

Figure 3.

!

CAUTION:

Do not over tighten.

Over tightening could break connector

and cause injury to operator.