Disposal, Troubleshooting – Heidolph LABOROTA 4000 eco User Manual

Page 64

Advertising
background image

64

E

CAUTION: avoid shocks and vibration while in transit.

DISPOSAL

For disposal, please comply with your local or national regulations.

Split by metal, plastic, etc.

Packing material to be treated as described above (material split).

TROUBLESHOOTING

No heat:

check power cord

power switch (1b) broken (call service engineer)

max. temperature limiter dropped (refer to Chapter “Adjust hot bath temperature”)

hot bath heater broken (call service engineer)

LABOROTA master switch (1b) OFF; (turn master switch 1b ON)

Drive won't run:

check power cord

master switch (1b) broken (call service engineer)

speed control knob at lefthand end, turn right

motor problem (call service engineer)

Insufficient vacuum:

check seals and fittings

check ground fittings, lubricate on condition

check PTFE seal for proper installation (sealing edge faces drive head). Refer to "set-
up, vapor tube and seal",

check vacuum pump's suction line

Advertising