Hmg 3010 – HYDAC HMG 3010 User Manual

Page 19

Advertising
background image

HMG 3010

Page 19

Edition 11/2011 - V04 R01

Part No.: 669855

The various options available for channels A
to H are demonstrated here using channel A
as an example. Select the measurement
range entry
on the right of the channel using
the navigation keys and press OK. This
takes you to the menu window for the se-
lected channel.
To enter a designation, press OK again
and then enter it in the box using the com-
bination keypad, followed by OK.
In the Mode menu you now have the option
of Automatic detection (HSI) and Manual.
The Automatic detection (HSI) option is
used for HYDAC sensors which have auto-
matic sensor detection. The Manual setting
is used for adapting sensors with standard
analogue signals (e.g. 4 - 20 mA) to the
HMG.

However, there are applications for which it
is useful to enter settings manually, also in
the case of HSI sensors with sensor recogni-
tion. For example, by entering settings
manually for a pressure measurement, you
can have physical variables displayed which
are proportional to the pressure, e.g. force.

Manual mode also provides the following
further settings:

Input signal
Upper and lower measurement range
Unit (of measurement)
Decimal format

As soon as you select one of these items, a
drop-down list appears or a box appears for
entering your settings.

Advertising