6.6 Technical data

Sensor inputs
The HMG features:

• 4 sensor input jacks (channel A – H) for up to 8 analogue sensors,

or up to 4 SMART sensors and

• 1 input jack with 2 digital inputs (channel I - J) and

a voltage input of -10 V to + 10 V (shown at channel H)

The sensors are connected using an standard M12x1 male connector (5 pole).

Channel A to H:

• Automatic detection for HSI sensors (pressure, temperature, flow rate trans-

mitters and SMART sensors)

• Connection of commonly-available sensors with current or voltage signals

• Differential channels for channel A - B; channel C - D; Differential channel for

flow rate measurement orifice (shown at channel B)

Channel I and J:

• Frequency channels (speed (rpm) measurement, counting function)

Frequency range: 1 .. 30,000 Hz

(≤ ± 0.1% FS max.)

Switching threshold / switch-back threshold: 2V/1V
Max. input voltage: 50 V; display delay max. 2 s.

Overview of signals / measurement inputs

Measurement

range

(accuracy)

Chann

.

A

Chann

.

B

Chann

.

C

Chann

.

D

Chann

.

E

Chann

.

F

Chann

.

G

Chann

.

H

HSI

X X X X X X X X

4 .. 20 mA

(

≤ ± 0.1 % FS max.)

X X X X X X X X

0 .. 20 mA

(

≤ ± 0.1 % FS max.)

X X X X X X X X

0 .. 4.5 V

(

≤ ± 0.1 % FS max.)

X X X X X X X X

0 .. 5 V

(

≤ ± 0.2 % FS max.)

X X X X X X

(

≤ ± 1.0 % FS max.)

X

X

0 .. 10 V

(

≤ ± 0.1 % FS max.)

X X X X X X

(

≤ ± 0.5 % FS max.)

X

X

0 .. 50 V

(

≤ ± 0.1 % FS max.)

X

X

0.5 .. 4.5 V

(

≤ ± 0.1 % FS max.)

X X X X x x x x

0.5 .. 5.5 V

(

≤ ± 0.1 % FS max.)

X X X X X X

(

≤ ± 1.0 % FS max.)

X

X

1 .. 5 V

(

≤ ± 0.2 % FS max.)

X X X X X X

(

≤ ± 1.0 % FS max.)

X

X

1 .. 6 V

(

≤ ± 0.2 % FS max.)

X X X X X X

(

≤ ± 0.5 % FS max.)

X

X

-10 .. +10 V

(

≤ ± 0.5 % FS max.)

X