Function, Technical data 4.3.1, Mechanical data – INFICON LDS3000 Modular Leak Detector User Manual

Page 13: Of application (see, Chapter 4.3, 2 function, 3 technical data, 1 mechanical data

Advertising
background image

Description

13

12 - STATUS

Status LED

The Power LED and Status LED indicate the status of the unit.

13 - CONTROL UNIT

Connection for control unit or I/O or bus module

Information The connections C

ONTROL

U

NIT

and I/O / A

NYBUS

have the same

functions. You have the choice of connecting:

– Control unit CU1000 + I/O module IO1000 or
– Control unit CU1000 + bus module BM1000 or
– I/O module IO1000 + bus module BM1000 or
– 2 control units CU1000 or
– 2 I/O modules IO1000 (configurable only together)

4.2

Function

The mass spectrometer module is a detection device for the test gases helium and
hydrogen. Integrated in test systems, the unit is used to detect gas being emitted
from a test object in order to indicate leaks.

The unit can be used both as a vacuum leak detector and a sniffer leak detector.

The MSB box outputs data on digital interfaces to the control unit CU1000, I/O
module IO1000 or bus module BM1000.

4.3

Technical data

4.3.1

Mechanical data

Power LED

Status LED

Meaning

Off

Red

Device not ready for operation

Green

Blue

Turbo molecular pump is starting

Green

Orange

Emission is switched on

Green

Green

Emission is stable

Green

Violet

Speed of the turbo molecular pump is
not within the normal range

Green

Error codes of the
status LED

Different activities of the unit

Green, flashes slowly

Supply voltage < 21.6 V

Green, flashes fast

Supply voltage > 26.4 V

Green, flashes fast

Off

Software is being updated

Green

Green, flashes fast

Software is being updated

Dimensions (W x H x D)

320 mm x 280 mm x 240 mm

Weight

14.3 kg

Connection G

ROSS

/F

OREPUMP

2 x DN 16

Connection F

INE

/S

NIFFER

2 x DN 16

Connection U

LTRA

DN 16 and DN 25

Advertising