Factory settings – INFICON LDS3000 Modular Leak Detector User Manual

Page 15

Advertising
background image

Description

15

4.3.5

Factory settings

Parameter

Factory setting

Operation mode

Vacuum

Bus module address

126

Pressure capillary surveillance (min.) 0.4 mbar

Pressure capillary surveillance (max.) 2 mbar

Pressure unit

mbar

Emission

On

Upper limit exponent

1 x 10

-5

Filter leak rate threshold

1 x 10

-10

Filter zero time

5 s

Filter mode

I•CAL

Gas ballast

Off

I/O module protocol

ASCII

Calibration request

Off

Calibr. factor xxx Mx

1

Calibration factor VAC/SNIF Mx

1.0
(for vacuum, sniffer and all masses)

Cathode selection

Auto Cat1

Compatibility mode

LDS3000

Config. Analog output 1

Leak rate mantissa

Config. Analog output 2

Leak rate exponent

Config. Analog output scale

0.5 V / decade

Configuration dig. outputs

Pin 1: Trigger 1, inverted
Pin 2: Trigger 2, inverted
Pin 3: Trigger 3, inverted
Pin 4: Trigger 4, inverted
Pin 5: Ready
Pin 6: Error, inverted
Pin 7: CAL request, inverted
Pin 8: Open, inverted

Configuration dig. Inputs

Pin 1: Select dyn. / normal CAL
Pin 2: Sniff
Pin 3: Start/Stop, inverted
Pin 4: Zero
Pin 5: External CAL
Pin 6: Internal CAL
Pin 7: Clear
Pin 8: -
Pin 9: -
Pin 10: -

Leak rate unit SNIFF

mbar*l/s

Leak rate unit VAC

mbar*l/s