Activate/deactivate zero, Selecting a signal filter, 8 activate/deactivate z – INFICON LDS3000 Modular Leak Detector User Manual

Page 31: 9 selecting a signal filter

Advertising
background image

Operation

31

6.8

Activate/deactivate Z

ERO

Z

ERO

can be used to suppress undesired helium backgrounds. If Z

ERO

is enabled,

the currently measured leakage rate value will be interpreted as a helium background
and subtracted from all subsequently measured values.

The background value suppressed by Z

ERO

is adjusted automatically if the

background changes inside the unit.

Information If the signal filter Fixed or 2-Zone is selected, then the background value

is automatically adjusted based on the configured zero time.

► Enable or disable Z

ERO

as needed:

S

ETTINGS

> Z

ERO

/F

ILTER

> Z

ERO

> Z

ERO

MODE

Z

ERO

WITH

S

TART

suppresses the helium background automatically when a

measurement is started.

► Enable or disable Z

ERO

WITH

S

TART

as needed:

S

ETTINGS

> Z

ERO

/F

ILTER

> Z

ERO

> Z

ERO

WITH

S

TART

6.9

Selecting a signal filter

► As a rule, use signal filter I•CAL.

► If the signal filter should simulate the time behavior of older units, then use filter

Fixed or 2-Zone.

► Select desired signal filter:

S

ETTINGS

> Z

ERO

/F

ILTER

> F

ILTER

> F

ILTER

MODE

Signal filter

I•CAL

The leakage rates are averaged at time intervals that are optimized
for the range of the leakage rates.

Fixed

The leak rates are averaged at fixed intervals of 0.2 seconds.

2-Zone

Filter compatible with LDS1000 and LDS2000
The averaging period is switched depending on the leak rate
threshold.

Advertising