1 error codes of the status led – INFICON LDS3000 Modular Leak Detector User Manual

Page 42

Advertising
background image

42

Operation

6.20.1 Error codes of the status LED

Any errors or warnings occurring in the MSB box will be indicated both as an error
code by the control unit and as a blink code by the Status LED.

The blink code is preceded by a long white signal. This is followed by an error or
warning number. An error number is indicated by means of red signals, while a
warning number is displayed using orange signals (the orange signals have a strong
green tinge, however):

• Start of blink code: long white signal

• Hundreds digit: 0 ... 9 red signals for errors / 0 ... 9 orange signals for warnings

• Break: blue signal

• Tens digit: 0 ... 9 red signals for errors / 0 ... 9 orange signals for warnings

• Break: blue signal

• Units digit: 0 ... 9 red signals for errors / 0 ... 9 orange signals for warnings

The blink code is repeated cyclically.

For example:

The pressure is too high.
-> Error code = Warning 520
-> Blink code of the Status LED: White (long), 5·orange, blue, 2·orange, blue