INFICON LDS3000 Modular Leak Detector User Manual
Page 48

48
Maintenance
Inserting a new operating fluid reservoir
The new operating fluid reservoir will be positioned correctly when you screw in the
cover:
1
Do not slide in the new operating fluid reservoir into the pump completely, but stop
once you reach the o-ring.
2
Use a face pin wrench to screw in the cover.
3
Insert a new o-ring.
4
Screw in the cover (tightening torque 13 Nm +/-10%).
5
Tighten the ventilation screw by hand.
6
Install the turbo molecular pump.
7
Put the mass spectrometer module into operation.
Confirm maintenance work
Control unit installed
Access = Integrator
► Confirm maintenance work on control unit:
A
CCESS
C
TRL
> I
NTEGRATOR
> M
AINTENANCE
> M
AINTENANCE
WORK
Material damage if o-ring is mounted improperly
An improperly mounted o-ring can cause leaks. The unit will experience
malfunctions and become damaged.
► Insert the o-ring carefully.