Operation

6.2.1

Factory layout

Analog output 1: Leak rate mantissa

Analog output 2: Leak rate exponent

6.2.2

Possible layouts

Off

The analog outputs are disabled (output voltage = 0 V).

Pressure p1 / Pressure p2

1 ... 10 V; 0.5 V / decade; 1 V = 1 x 10

-3

mbar

Leak rate mantissa

1 ... 10 V; linear; in used unit

Information Useful only if the other analog output is assigned “Leak rate exponent”.

Leak rate exponent

1 ... 10 V; 0.5 V / decade; step function; 1 V = 1 x 10

-12

; in used unit

Information Useful only if the other analog output is assigned “Leak rate mantissa”

or “Leak rate ma. hys.”.

Leak rate linear

x ... 10 V; linear; in used unit

The upper limit (= 10 V) is set via the parameter “Upper limit exponent”.

Leak rate log.

x ... 10 V; logarithmic; in used unit

The upper limit (= 10 V) and the scale (V / decades) are set via the parameters
“Upper limit exponent” and “Scale for leak rate”.

For example:
Upper limit set to 10

-5

mbar·l/s (= 10 V).

Scale set to 5 V / decade.
Lower limit is 10

-3

mbar·l/s (= 0 V).