10 calibrating the unit – INFICON LDS3000 Modular Leak Detector User Manual

Page 32

Advertising
background image

32

Operation

6.10

Calibrating the unit

► To ensure measuring accuracy, the unit must be calibrated regularly.

Information Recalibration at the start of every shift is recommended.

Incorrect measurement results due to premature calibration

If calibrated prematurely, the unit will produce incorrect measurement results.
► Allow the unit to run for at least 20 minutes before performing the calibration.

Calibration

Particularities

internal

• with internal test leak

• Auto tune (mass adjustment)

• Determining the calibration

factor with the signal of the test
leak tuned

• Determining the background

Following calibration, adjust
the machine/sniff factor as
needed, see

Chapter 6.10.4,

page 34

external

• Vacuum mode: with external

test leak installed in the test
system

• Sniffer mode: with external

sniffer leak

• Taking the test system

characteristics into account
(pressure, split flow ratio)

• Auto tune (mass adjustment)

• Determining the calibration

factor after tuning the signal of
the test leak

• Determining the background

external -
dynamic

• with external test leak installed

in the test system

• Taking the test system

characteristics into account
(pressure, split flow ratio,
measuring time)

• Measuring time based on the

dynamic signal curve

• Determining the calibration

factor before tuning the signal
of the test leak

• Determining the background

Advertising