INFICON LDS3000 Modular Leak Detector User Manual

Page 41

Advertising
background image

Operation

41

* compatib. LDS1000/LDS2010 or binary protocol

611

W

Machine factor too high,
(machine factor)

81

153

1.00E+
04

Machine factor adjustment
inaccurate, split flow factor
too high

612

W

Machine factor lower
than last,
(machine factor)

81

153

< 50%
old
value

Split flow factor has
changed

613

W

Machine factor higher
than last,
(machine factor)

81

153

> 200%
old
value

Split flow factor has
changed

625

W

Int testleak not set,
(int. Testleak)

0

0

Leakage rate of internal
test leak is still set to
factory setting

626

W

Ext testleak not set,
(ext. Testleak)

0

0

Leakage rate of test leak is
still set to factory setting

630

W

Calibration request

0

0

Temperature change of
5°C, speed was changed
since last calibration, 30-
minute switch-on time and
still no calibration
conducted

7xx temperature errors (pre-amplifier, electronics)

700

W

Preamplifier temp too
low,
(temperature [°C])

33

60

2°C

Temperature too low

702

W

Preamplifier temp too
high,
(temperature [°C])

32

124

60°C

Temperature too high

703

E

Preamplifier not
connected

33

60

Pre-amplifier defective,
cable not plugged in

710

W

Max. MSB temperature
exceeded,
(temperature [°C])

54

44

55°C

Temperature too high

711

E

MSB temperature far
too high,
(temperature [°C])

54

44

65°C

Temperature too high

8xx not used

9xx Maintenance warnings (e. g. TMP)

901

W

Maintenance bearing/
lubricant

99

99

3 years

TMP maintenance
necessary

Error

number

Warning/

Error

Error message
LDS3000

Error

number

LDS1000

protocol

Error

number

ASCII

protocol*

Limit
values

Cause