Crystal12 Sensor Operating Manual

Chapter 1

Sensor Specifications

1.1 Specifications for Crystal12 Sensor 750-667-G1

The Crystal12 sensor is a twelve crystal quartz crystal deposition monitor
transducer. When the monitor crystal being used fails, a new crystal is advanced
into position, fully replacing the failed crystal’s function. The mechanism used for
providing this motion is pneumatically powered. The logic for automatic operation
of this transducer is provided with the Cygnus™ Deposition Controller.
See

Figure 1-1

and

Table 1-1

.

Figure 1-1 Crystal12 Sensor

Table 1-1 Sensor Specification

Maximum bakeout temp with no water

130 °C

Maximum operating isothermal
environment temperature with minimum
water flow

300 °C

Size (maximum envelope)

4.0" (102 mm) dia. x 3.3" (84 mm) high
4.75" (121 mm) dia. x 3.46" (88 mm) with
optional mounting posts installed

Water, air and coax length

Standard 30" (762 mm)

Crystal exchange

Front-loading

Mounting

Six #4-40 tapped holes on the back of the
sensor body, six #4-40 tapped holes on
outside circumference. Three #6-32
tapped holes with optional mounting kit.