Crystal12 Sensor Operating Manual

1.1.1 Installation Requirements

1.1.1.1 Feedthrough

Š Qty (1) 2-3/4" ConFlat® with two coaxial feedthroughs, two pass water, one air

IPN 002-080, See

Figure 5-2 on page 5-2

or,

Š Qty (1) 1" bolt with 1 coaxial feedthrough, two pass water, one air IPN

750-030-G1, see

Figure 5-1 on page 5-1

.

1.1.1.2 Other

Š User to provide mounting structure adequate to support weight of Crystal12

and designed to facilitate removal and replacement with minimal change in
exact position. An optional mounting post kit, IPN 750-670-G1, may be
purchased for this purpose.

CAUTION

Do not mount the Crystal12 unsupported.

Š User to provide vacuum-tight braze joints or connectors for the water and air

tubes.

Š Valve assembly for air, IPN 750-420-G1 (not provided), with a 0.022" restrictor

orifice installed by the user. (Orifice included with Crystal12 accessory kit.) See

Table 1-3 on page 1-4

.

Š XIU or Oscillator designed to interface with the Cygnus controller.

1.1.1.3 Utilities

Š Minimum water flow 150-200 cc/min, 30 °C max (Do not allow water to freeze).

Coolant should not contain chlorides as stress corrosion cracking may occur.

Š Regulated air supply 80-90 PSIG (5.5 bar - 6.2 bar) [550 kPa - 620 kPa]

Š 2 meter maximum length of 1/8" tubing between bellows assembly and the

control valve.

CAUTION

Do not allow water tubes to freeze. This may happen if the
tubes pass through a cryogenic shroud and the fluid’s
flow is interrupted.