4 - 11

07

4-

39

8-

P1A

Crystal12 Sensor Operating Manual

While in Ready or Stop,
initiating a Crystal Switch
function causes unit to
index more than once.

Resistor is open at
bypassed position(s).

Repair resistor at
bypassed position of
750-652-G1 (item 15 of

Figure 1-5 on page 1-8

).

Or, contact INFICON.

Unit indexes twelve times
and displays Xtal Switch
Error, Xtal Fail, and
Carousel Change-Out
message.

Loss of electrical signal

Check for electrical
continuity and isolation.

Resistor #1 is open /
shorted.

No carousel installed

Install carousel

Torsion spring of electrical
connection assembly
broken.

Replace electrical
connection assembly.
Refer to

section 4.3.4 on

page 4-7

Table 4-1 Troubleshooting (continued)

Symptom

Cause

Remedy

Advertising