6

Actuate the bellows assembly by applying 24 volts to the leads of the solenoid
valve of the solenoid valve assembly. Consult

Figure 3-3 on page 3-5

for wiring

assignments. The air pressure must be sustained.

7

Rotate the ratchet (item 29,

Figure 1-5 on page 1-8

) counterclockwise until one

tooth contacts the pin of the pawl and actuator stem assembly (item 18,

Figure

1-4 on page 1-7

). Secure the ratchet to the 0.25" diameter shaft of the carousel

spindle body (item 24,

Figure 1-5 on page 1-8

) utilizing the #6-32 x 0.125" long

socket set screw (item 12,

Figure 1-4 on page 1-7

).

8

Rotate the stop ratchet (item 31,

Figure 1-5 on page 1-8

) clockwise until it

contacts the stem of the pawl and actuator stem assembly (item 18,

Figure 1-4

on page 1-7

) as shown in

Figure 4-2

. Secure the stop ratchet to the 0.25”

diameter shaft of the carousel spindle body (item 24,

Figure 1-5 on page 1-8

)

utilizing the #6-32 x 0.125 long socket set screws (item 12,

Figure 1-5 on page

1-8

).

Figure 4-2 Setting the Stop Ratchet

9

Position the detent spacing tool (of the 750-254-G2 assembly) between the
detent (item 25,

Figure 1-5 on page 1-8

) and ratchet (item 29,

Figure 1-5 on

page 1-8

) as shown in

Figure 4-3

. Pull the detent against the detent spacing

tool and subsequently against the tooth of the ratchet. Tighten the #2 fastener
(item 5,

Figure 1-5 on page 1-8

) to secure the detent to the bellow assembly.