4.4 Measurement mode

4.4.1

Selection

The Vacuum Gauge Controller automatically selects the
measurement mode after it has been switched on.

When the unit is set to the parameter mode, it will auto-
matically return to the measurement mode if no button is
pressed for 10 seconds.

4.4.2

Description

The measurement mode is the standard operating mode.
It displays the pressure readings of the sensors. A status
message is displayed if the pressure exceeds the permis-
sible range. See Tab. 4-1,  17.

Tab. 4-1

Display when in measurement mode

Channels which are not connected to a sensor display
noSEn

. This status message disappears after approxi-

mately two minutes.

4.4.3

Control button functions

4.4.3.1

Selecting a channel

1

Press the CH button

Fig. 4-2

Pressing the CH button

The unit selects the next channel. The number of the
selected channel is flashing for a few seconds.

4.4.3.2

Selecting parameter mode

1

Press the PARA button

Fig. 4-3

Pressing the PARA button

The unit changes to the parameter mode. See
Chapter 4.5 Parameter mode,  19. It will automatically
return to the measurement mode if no button is pressed
for 10 seconds.

4.4.3.3

Switching high-vacuum circuit on

The high-vacuum circuit of the following sensors can be
switched on manually: PEG.

For this purpose the sensor control must be set to HAnd.
See Chapter 5.2.6 Sensor activation (S-on),  25.

1

Press the CH button to select the required channel

2

Keep the UP button pressed for approximately 1
second

Fig. 4-4

Press the UP button for 1 second

The sensor on the selected channel is switched on. The
HV indicator is illuminated. The display shows the pres-
sure reading or a status message. See Tab. 4-1,  17.

Display

Pressure

Er Hi

Significantly above the permissible
range

The FAIL indicator flashes
The error signal relay switches

Reading

In the permissible range

Er Lo

Significantly below the permissible
range

The FAIL indicator flashes
The error signal relay switches

Er x

Error message of BPG400/HPG
x = Error code (High-Byte)

Er xx

Error message of BCG, BPG402
xxH = Error code

noSEn

See Tab. 4-2,  18

noid

See Tab. 4-2,  18

oFF

See Tab. 5-14,  30

Hot

See Chapter 5.2.6,  25

SELF

See Chapter 5.2.8,  25

CH 1

See Chapter 5.2.8,  25

CH 2

See Chapter 5.2.8,  25

CH 3

See Chapter 5.2.8,  25

LoC

See Chapter 5.4.5,  30

para

ch

para

ch

para

ch