10
tina95e1 (2014-07)
2 Technical
Data
Measurement range
→ "Validity"
Accuracy
1)
0.1 … 0.5 F.S.
0.1 … 1100 F.S.
0.4% of reading
0.2% of reading
Temperature effect on zero
0.1 … 0.5 F.S.
1 … 1100 F.S.
0.0050% F.S./ °C
0.0025% F.S./ °C
Temperature effect on span
0.02% of reading / °C
Resolution
0.003% F.S.
Gas type dependence
none
Output signal analog
(measurement signal)
Measurement range
0 … +10 V
Voltage range
–5 … +10.5 V
(limited to +10.5 V)
Relationship voltage-pressure
linear
Output impedance
0
Ω (short-circuit proof)
Loaded impedance
>10 k
Ω
Step response time
2)
Analog output signal
Digital output signal (EtherCAT
interface)
≤1 ms
≤2 ms
Identification
Resistance R
Ident
Voltage
13.2 k
Ω referenced to sup-
ply common
≤5 V
1)
Non-linearity, hysteresis, repeatability at 25 °C ambient operating
temperature without temperature effects after operation of 2 h.
2
)
Increase 10 … 90 % F.S.R.