INFICON Stripe CDG100Dhs 0.1 ... 1000Torr / mbar User Manual
Page 12
12
tina95e1 (2014-07)
EtherCAT connector
2×RJ45, 8-pin, socket
input and output
Cable 8-pin,
shielded,
Ethernet
Patch Cable (CAT5e quality
or higher)
Cable length
≤100 m
For further information on the EtherCAT interface
→ [3]
Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (PELV). The connection to the gauge
has to be fused
4)
.
Supply voltage
at the gauge
+14 … +30 V (dc) or
±15 V (dc)
Power consumption
while being heated
at
operating
temperature
≤16 W
≤11 W
Fuse to be connected
4)
≤1.25 AT
The gauge is protected against reverse polarity of the supply
voltage and overload.
Electrical connection
15-pin D-Sub, male
Sensor cable
15-pin plus shielding
4)
INFICON controllers fulfill this requirement.