INFICON Stripe CDG100Dhs 0.1 ... 1000Torr / mbar User Manual

Page 13

Advertising
background image

tina95e1 (2014-07)

13

Cable length

Supply voltage 15 V


≤ 8 m (0.14 mm²/conductor)
≤15 m (0.25 mm²/conductor)

Supply voltage 24 V

≤43 m (0.14 mm²/conductor)
≤75 m (0.25 mm²/conductor)

Supply voltage 30 V

≤88 m (0.14 mm²/conductor)
≤135 m (0.25 mm²/conductor)

For longer cables, larger conductor cross-sections are required
(R

cable

≤1.0

Ω).

Grounding concept

→ "Power Connection"

Materials exposed to vacuum

ceramics (Al

2

O

3

≥99.5%),

stainless steel AISI 316L

Internal volume

≤4.2 cm

3

Admissible pressure (absolute)

1000 / 1100 F.S.
1 … 500 F.S.
0.1 … 0.5 F.S.


3 bar
2.6 bar
1.3 bar

Bursting pressure (absolute)

6 bar

Admissible temperatures

Storage
Operation
Bakeout

–20 °C … +85 °C
+10 °C … +50 °C
≤110 °C at the flange

Relative humidity

≤80% at temperatures
≤+31 °C, decreasing to 50%
at +40°C

Use

indoors only, altitude up to
3000 m NN

Degree of protection

IP 30

Mounting orientation

standing upright to horizon-
tal

Advertising