Briefly press the <ZERO> button with a pin (max.

ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration ≤8 s)
is completed.

max. ø1.1 mm

Press the
button briefly

After zero adjustment, the gauge automatically returns

to the measurement mode.

The <STATUS> LED blinks green if

• the signal output is negative (< -20 mV) when the final pres-

sure has been attained

• the zero adjustment has failed.

4.2.2

<ZERO> Adjustment with Ramp Function

The ramp function allows to adjust the pressure value at a
known reference pressure within the measurement range of the
gauge.

It also permits to adjust an offset of the characteristic curve in
order to
• compensate for the offset of the measuring system or
• obtain a slightly positive zero for a 0 … 10 V AD converter.
The offset should not exceed 5% of the F.S. (+500 mV). At a
higher positive offset, the upper limit of the measurement range
is exceeded.