6 activating the factory setting (factory reset) – INFICON Stripe CDG100Dhs 0.1 ... 1000Torr / mbar User Manual
Page 35
tina95e1 (2014-07)
35
The current ATM threshold setting
• can be read / written via the diagnostic port
• is output at the measurement signal output instead of the
pressure signal, can be measured with a voltmeter, and is
displayed on the LCD display after the <SP1> or <SP2>
button is pressed
• can be read / written via the EtherCAT interface.
DANGER
DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
malfunction will cause.
Programming ATM setpoint
Programmable parameters:
(
→ [3], [4])
Factor of ATM
Low Trip Enable
Low Trip Point Hysteresis
High
Trip
Enable
High Trip Point Hysteresis
Setpoint
Mode
Switching characteristics of the setpoints
→ 29.
Diagnostic port
→ [4].
4.6
Activating the Factory Setting (Factory Reset)
All user defined parameters (e.g. zero, filter) are restored to their
default values.