2 technical data – INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual

Page 10

Advertising
background image

10

tina86e1 (2013-11)

2 Technical

Data

For further technical data for gauges with EtherCAT

interface

→  [5].

Measurement range

→ "Validity"

Accuracy

1)

0.2% of measured value

Temperature effect on zero 0.0025%

F.S./

°C

Temperature effect on span

0.02% of reading / °C

Resolution

0.003% F.S.

Gas type dependence

none

Output signal analog
(measurement signal)

Measurement range

0 … +10 V

Voltage range

–5 … +10.24 V
(limited to +10.24 V)

Relationship voltage-pressure

linear

Output impedance

0

Ω (short-circuit proof)

Loaded impedance

>10 k

Ω

Response time

2)

30 ms

Identification

Resistance R

Ident


Voltage


13.2 k

Ω referenced to

supply common
≤5 V

Remote Zero Adjust


External switching contact

digital input for zero adjust-
ment (

→  25)

30 V (dc) / <5 mA (dc)

1)

Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C

ambient operating temperature without temperature effects after
operation of 2 h.

2)

Increase 10 … 90 % F.S.R.

Advertising
This manual is related to the following products: