21
If no sensor cable is available, make one according to the
following diagram (cable length and conductor cross-sections
→ 12).
case
15 V
15 V
14…30 V
Remote
Zero Adjust
1 M
4
1
8
9
10
Ident
13
14
2
TxD
RxD
12
7
11
5
15
100
nF
18 V
10
6
3
SP1
SP2
Status
Pin 1,4
Relay SP1, closing contact
Pin 2
Signal output (measurement signal)
or thresholds SP1/2
Pin 3
Status
Pin 5
Supply common
Pin 6
Supply (-15 V)
Pin 7, 11 Supply (+14 … +30 V or +15 V)
Pin 8, 9 Relay SP2, closing contact
Pin 10
Gauge identification
or Remote Zero Adjust
Pin 12
Signal common
Pin 13
RS232, TxD
Pin 14
RS232, RxD
Pin 15
Housing (chassis ground)
case Connector