INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual

Page 15

Advertising
background image

tina86e1 (2013-11)

15

Analog Measurement Signal vs. Pressure

1

2

3

4

5

6

7

8

9

10

0

0.9×F.S.

0.5×F.S.

0.6×F.S.

1.0×F.S.

0.7×F.S.

0.8×F.S.

0.1×F.S.

0.2×F.S.

0.3×F.S.

0.4×F.S.

1.1×F.S.

Pressure p

Measuring signal U

out

[V]

0.0×F.S.

p = (U

out

/ 10 V) × p (F.S.)

Conversion Torr

↔ Pascal

Torr

mbar

5)

Pa

4)

c 1.00

1013.25 / 760 =

1.3332…

101325 / 760 =

133.3224…


Example: Gauge with 10 Torr F.S.

Measurement signal U

out

= 6 V

p = (6 V / 10 V) × 10 Torr

= 0.6 × 10 Torr = 6 Torr

5)

Source: NPL (National Physical Laboratory)

Guide to the Measurement of Pressure and Vacuum, ISBN 0904457x /
1998

Advertising