2 power connection – INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual

Page 20

Advertising
background image

20

tina86e1 (2013-11)

3.2 Power

Connection

Make sure the vacuum connection is properly made

(

→  16).

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused

6)

.

Ground loops, differences of potential, or EMC problems

may affect the measurement signal. For optimum signal
quality, please do observe the following notes:

• Use an overall metal braided shielded cable. The

connector must have a metal case.

• Connect the cable shield to ground at one side via the

connector case. Make sure the connector case has
direct contact to the cable's shield on its whole cir-
cumference.

Do not connect the other side of the

shield.

• Connect the supply common with protective ground

directly at the power.

• Use differential measurement input (signal common

and supply common conducted separately).

• Potential difference between supply common and

housing ≤18 V (overvoltage protection).

6)

INFICON controllers fulfill this requirement.