Supply

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused

3)

.

Supply voltage

at the gauge

Class 2 / LPS
+14 … +30 V (dc) or
±15 V (±5%)

Ripple

≤1 V

pp

Power consumption

while being heated
CDG045D2
CDG100D2



≤15 W
≤20 W

at operating temperature

CDG045D2
CDG100D2


≤10 W
≤14 W

The gauge is protected against reverse polarity of the supply
voltage and overload.

Electrical connection

D-Sub 15-pin, male

Sensor cable

15-pin plus shielding

Cable length

Supply voltage 15 V


≤ 4 m (0.14 mm²/conductor)
≤ 7 m (0.25 mm²/conductor)

Supply voltage 24 V

≤20 m (0.14 mm²/conductor)
≤35 m (0.25 mm²/conductor)

Supply voltage 30 V

≤40 m (0.14 mm²/conductor)
≤70 m (0.25 mm²/conductor)

For longer cables, larger conductor cross-sections are required
(R

cable

≤1.0

Ω).

3)

INFICON controllers fulfill this requirement.