Push the <ZERO> button again:

Fine adjustment
within 0...3 s:

the zero adjustment value changes
by one unit

Change of direction
within 3...5 s:

the zero adjustment changes its
direction (the blinking frequency of
the <STATUS> LED changes
briefly)

If the <ZERO> button is released for more than 5 s, the

gauge returns the measurement mode.

The upper threshold is automatically set 1% F.S. above

the lower one (hysteresis).

Adjusting Setpoint <2>

Push the <SP> button twice (the LED <2> blinks). The
adjustment procedure is the same as for setpoint <1>.

4.4

Activating the Factory Setting (Factory Reset)

All user defined parameters (e.g. zero, filter) are restored to their
default values.

Loading of the default parameters is irreversible.

Loading the default parameters:

Put the gauge out of operation.

Keep the <ZERO> button depressed for at least 5 s while

the gauge is being put into operation (Power ON).