INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual

Page 27

Advertising
background image

tina86e1 (2013-11)

27

It also permits to adjust an offset of the characteristic curve in
order to
• compensate for the offset of the measuring system or
• obtain a slightly positive zero for a 0 … 10 V AD converter.
The offset should not exceed 2% of the F.S. (+200 mV). At a
higher positive offset, the upper limit of the measurement range
is exceeded.

Zero adjustment using the ramp function can be

performed via
• the <ZERO> button on the gauge,
• the diagnostic port (→  [4]),
• the RS232C interface (→  [4]).

Recommended procedure for adjusting the offset of a

measuring system:

→ Notice  24.

Operate the gauge for at least 90 minutes under constant

ambient conditions (until the signal is stable).

Push the <ZERO> button with a pin (max. ø1.1 mm) and

keep it depressed. The <STATUS> LED starts blinking.
After 5 s, the zero adjustment value, starting at the current
output value, keeps continually changing (ramp) until the
button is released or until the setting limit (max. 25% F.S.)
is reached. The corresponding output signal is delayed by
about 1 s.