3 switching functions – INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual

Page 29

Advertising
background image

tina86e1 (2013-11)

29

4.3 Switching

Functions

The two switching functions can be set to any pressure within
the measurement range of the gauge (

→  15).

The current setpoint setting
• can be read/written via the diagnostic port,
• is output at the D-Sub connector instead of the measurement

signal (

→  21) and can be measured with a voltmeter after

the <SP> button is pressed, or

• can be read/written via the RS232C interface.
If the pressure is lower than the setpoint, the corresponding LED
(<SP1> or <SP2>) is lit solid and the corresponding relay
(

→  21) is energized.

Measurement signal (pressure p)

Time t

Me

asu

reme

nt value

Off

On

Hysteresis
(1% F.S.)

Off

Threshold value

Setpoint

Advertising
This manual is related to the following products: