INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual

Page 10

Advertising
background image

10

tina03e1-b (2004-02) BPG400 v1.om

Data rate
Data format

Connections (sensor cable connector)
TxD (Transmit Data)
RxD (Receive Data)
GND

9600 Baud
binary
8 data bits
one stop bit
no parity bit
no handshake

Pin 13
Pin 14
Pin 5

(Function and communication protocol of the RS232C interface

→ 31)

Fieldbus name

DeviceNet

Standard applied

→ [6]

Communication protocol, data format

→ [1], [4]

Interface, physical

CAN bus

Data rate (adjustable via "RATE"
switch)

125 kBaud
250 kBaud
500 kBaud
"P" (125 kBaud, 250 kBaud, 500 kBaud
programmable via DeviceNet
(

→ [1])

Node address (MAC ID)
(Adjustable via "ADDRESS", "MSD",
"LSD" switches)

0 … 63

dec

"P" (0 … 63 programmable via
DeviceNet,

→ [1])

DeviceNet connector

Micro-Style, 5 pins, male

Cable

Shielded, special DeviceNet cable,
5 conductors (

→ 22 and [4])

Cable length, system wiring

According to DeviceNet specifications
(

→ [6], [4])

Fieldbus name

Profibus

Standard applied

→ [7]

Communication protocol data format

→ [10], [7]

Interface, physical

RS485

Data rate

≤12 MBaud (→ [10])

Node address

Local
(Adjustable via hexadecimal
"ADDRESS", "MSD", "LSD"
switches)
Via Profibus
(hexadecimal "ADDRESS" switches
set to >7d

hex

(>125

dec

))

00 … 7D

hex

(0 … 125

dec

)

00 … 7D

hex

(0 … 125

dec

)

Profibus connection

D-Sub, 9 pins, female

Cable

Shielded, special Profibus cable
(

→ 23 and [5])

Cable length, system wiring

According to Profibus specifications
(

→ [7], [5])

RS232C interface

DeviceNet interface
(BPG400-SD)

Profibus interface
(BPG400-SP)

Advertising