For reasons of compatibility, the expression "sensor cable" is used for all
BPG400 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BPG400-SD, BPG400-SP or
BPG400-SR) is normally transmitted via DeviceNet, Profibus or RS485.
The sensor cable is required for supplying all BPG400 types with power.
In connection with the gauges with fieldbus interface (BPG400-SD,
BPG400-SP and BPG400-SR), it also permits access to the relay con-
tacts of the switching functions (

→ 21, 44).

The application and length of the sensor cable have to be considered when deter-
mining the number and cross sections of the conductors (

→ 9).

Open the cable connector (D-Sub, 15 pins, female).

Prepare the cable and solder/crimp it to the connector as indicated in the
diagram of the gauge used:

Electrical connection
Pin 2 Signal output (measuring signal)
Pin 5 Supply common, GND
Pin 7 Degas on, active high
Pin 8 Supply
Pin 10 Gauge identification
Pin 12 Signal common, GND
Pin 13 RS232C, TxD
Pin 14 RS232C, RxD
Pin 15 Shielding, housing, GND

8

9

1

15

+24 VDC
+24 VDC

Pins 1, 3, 4, 6, 9 and 11 are
not connected internally.

0 … +10 V

D-Sub, 15 pins

female,

soldering side

TxD

BPG400

RxD

Measuring

signal

Degas

+Ub

13
14

2

12

7

8
5

15

+

1.25AT

Identification

10

RS232C

+

24 V

3.2.2.1 Making an Individual

Sensor Cable

Cable type

Procedure

Sensor cable connection
BPG400