3 syntax description – INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual

Page 39

Advertising
background image

tina03e1-b (2004-02) BPG400 v1.om

39

In the table "Commands and Responses" (

→ 40) the following variables are

used:

Variable Description

Values, range,

aa

Operating device
address

00 … 3F

hex

fff

Emission current

_25UA
5.0MA
_20MA

= 25

µA

= 5 mA
= 20 mA

modeR

Gauge operation
mode, response

BPG_400_
RIG_

_

_

_

_

= BPG400 mode
= RIG mode (reduced ion

gauge mode)

modeT

Gauge operation
mode, command

BPG400
RIG

= BPG400 mode
= RIG mode (reduced ion

gauge mode)

n

Filament selected

1
2

No action, for future use

oo

Address offset

00 … 3F

hex

rate

Data transfer rate

300, 1200, 2400, 4800,
9600, 19'200 or 28'800 Baud

status

Gauge status

BPG mode:
BPG_ST_0
BPG_ST_5
BPG_ST_8
BPG_ST_9
RIG mode:
01_OVPRS
02_EMISS
00_ST_OK

= no error
= Pirani warning
= BA error
= Pirani error

If emission is switched off
= BA error
= normal operation

t

Toggle status

ON
OFF

= toggle function on
= toggle function off

unit

Pressure unit

MBAR, TORR, PASCAL

v.vv

Firmware version of
gauge

Three characters and a decimal point. Higher
revision numbers indicate newer versions of
firmware. Example: 1.00.

x.xx

Mantissa of
pressure values

1.00 … 9.99

syy

Signed exponent of
pressure values

s

yy

= Sign of exponent, +/

= Exponent

00 … 09

z

Operation of relays
(switching function)

+

= Relay on below threshold
= Relay on above threshold

All Commands and responses are terminated by "carriage return" (CR).

All response messages contain 13 characters, including CR.

For better readability, a space character is signified by a " _ " (underline)
character in the table below.

4.9.3 Syntax Description

4.9.3.1 Definitions, Legend