The programming procedure of the switching functions via the RS485 interface im-
plemented in the BPG400-SR differs from the one used on BPG400-SD/SP.

The table below describes the programming possibilities.

Function

Threshold values,
relay operation

Programming
example
(

→ table, 40)

Mea

sured

value

Off

Off

On

Hysteresis

Re

la

y

Pressure p

SL -

SL +

SL+ < SL–
⇒ relay on below SL+

T: #02SL+1.00E–04 CR
R: *02_PROGM_OK CR
T: #02SL–2.00E–04 CR
R: *02_PROGM_OK CR

Mea

sured

value

On

On

Off

Hysteresis

Rel

ay

Pressure p

SL +

SL -

SL+ > SL–
⇒ relay off below SL–

T: #02SL+2.00E–04 CR
R: *02_PROGM_OK CR
T: #02SL–1.00E–04 CR
R: *02_PROGM_OK CR

Mea

sured

value

Off

Off

On

Minimal

Hysteresis

Rel

ay

Pressure p

SL + = SL -

2

1

SL– = SL+,
SL+ command sent last
⇒ relay off below
SL+/SL–
(minimal hysteresis)

T: #02SL–3.00E–04 CR
R: *02_PROGM_OK CR
T: #02SL+3.00E–04 CR
R: *02_+MIN_HYS CR

Mea

sured

value

On

On

Off

Minimal

Hysteresis

Rel

ay

Pressure p

SL - = SL +

2

1

SL+ = SL–,
SL– command sent last
⇒ relay on below
SL+/SL–
(minimal hysteresis)

T: #02SL+3.00E–04 CR
R: *02_PROGM_OK CR
T: #02SL–3.00E–04 CR
R: *02_–MIN_HYS CR

Programming the second switching function (setpoint B) is identical to
the procedure described above. Threshold variables are SH+ and SH–
in this case.

4.9.4.1 Programming the

Switching Functions