4 switching functions – INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual

Page 42

Advertising
background image

42

tina03e1-b (2004-02) BPG400 v1.om

The BPG400-SR has two independent switching functions. A floating relay contact
is available for each switching function (

→ 21). Additionally, the relay contact of

the switching function A is accessible on the RS485 interface connector (

→ 24).

The functionality of the switching functions (setting of thresholds and re-
lay operation) depends on the gauges threshold potentiometer settings:

Threshold potentiometer setting Functionality of switching function

≤0.5 Volt

1)

• Thresholds defined by stored val-

ues (SL+/–, SH+/–, sent by the
host)

• Hysteresis defined by independent

setting of "+" and "–" values

• Relay operation defined by "+" and

"–" values and sequential order of
data transfer

(

→ following table)

>0.5 Volt

1)

• Thresholds defined by potentio-

meter settings

• Hysteresis = 10% of threshold
• Relays energize when pressure

falls below threshold

(Same as BPG400-SD/SP,

→ 44)

1)

Threshold voltages can be read by the host using the appropriate commands
(

→ 41) or measured on the corresponding pins of the gauges 25 pin D-Sub

connector (

→ 21).

4.9.4 Switching Functions

Selecting the Source of the
Threshold Values