3

Over the whole measuring range, the gauge has a continuous characteristic curve
and its measuring signal is output as logarithm of the pressure.
The gauge functions with a Bayard-Alpert hot cathode ionization measurement
system (for p < 2.0×10

-2

mbar) and a Pirani measurement system (for

p > 5.5×10

-3

mbar). In the overlapping pressure range of

2.0Ч10

-2

… 5.5Ч10

-3

mbar, a mixed signal of the two measurement systems is

output. The hot cathode is switched on by the Pirani measurement system only
below the switching threshold of 2.4×10

-2

mbar (to prevent filament burn-out). It is

switched off when the pressure exceeds 3.2×10

-2

mbar.

DeviceNet™