tina03e1-b (2004-02) BPG400 v1.om

Plug the sensor connector into
the gauge and secure it with the
locking screws.

Connect the other end of the sensor cable to the connector of the instru-
ment or gauge controller you are using.

The gauge can now be operated via analog and RS232C interface.

For operating BPG400-SD via DeviceNet, an interface cable conforming to the
DeviceNet standard is required.
If no such cable is available, make one according to the following indications.

A shielded special 5 conductor cable conforming to the DeviceNet standard has to
be used (

→ [4], [6]).

Make the DeviceNet cable according to the following indications.

3

1

4

2

5

Micro-Style, 5 pins,
(DeviceNet)
female, soldering side

Pin

Function (BPG400-SD)

1

Drain

2

Supply

+24 VDC

(DeviceNet) interface only

3

Supply common GND

(DeviceNet interface only)

4

CAN_H

5

CAN_L