6 rs232c interface – INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual

Page 31

Advertising
background image

tina03e1-b (2004-02) BPG400 v1.om

31

The built-in RS232C interface allows transmission of digital measurement data and
instrument conditions as well as the setting of instrument parameters.

Caution

Caution: data transmission errors
If the gauge is operated with the RS232C interface and a fieldbus in-
terface at the same time, data transmission errors may occur.
The gauge must not be operated with the RS232C interface and
DeviceNet, Profibus or RS485 at the same time.

The interface works in duplex mode. A nine byte string is sent continuously without
a request approx. every 20 ms.
Commands are transmitted to the gauge in a five byte input (receive) string.

• Data rate

9600 Baud

set value, no handshake

• Byte

8 data bits
1 stop bit

• TxD

Pin 13

• RxD

Pin 14

• GND

Pin 5

(Sensor cable connector)

The complete output string (frame) is nine bytes (byte 0 … 8). The data string is
seven bytes (byte 1 … 7).

Byte No Function

Value

Comment

0

Length of data string

7

(Set value)

1

Page number

5

(For BPG400)

2

Status

→ Status byte

3

Error

→ Error byte

4

Measurement high byte

0 … 255

→ Calculation of pressure value

5

Measurement low byte

0 … 255

→ Calculation of pressure value

6

Software version

0 … 255

→ Software version

7

Sensor type

10

(For BPG400)

8

Check sum

0 … 255

→ Synchronization

Synchronization of the master is achieved by testing three bytes:

Byte No Function

Value

Comment

0

Length of data string

7

Set value

1

Page number

5

(For BPG400)

8

Check sum of bytes No 1 … 7

0 … 255

Low byte of check
sum

1)

1)

High order bytes are ignored in the check sum.

4.6 RS232C Interface

4.6.1 Description of the

Functions

Operational parameters

Electrical connections

4.6.1.1 Output String (Transmit

)

Format of the output string

Synchronization