Page 46: 6 commissioning/start-up

6 Commissioning/start-up
46
6.6
Setting the filter time constant (damping) (DamP)
The filter time constants (damping) can be used to "damp" the measured value.
Small filter time constant: the measured value is updated faster.
Large filter time constant: the measured value is updated slower.
The value is entered in seconds with two decimal places.
How to proceed:
6.7
Setting the mains frequency (Freq)
This setting can be used to change the mains frequency from 50 Hz to 60 Hz.
How to proceed:
Step
Activity
1
Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2
"Turn" until "DamP" appears in the bottom line.
3
"Press"
Step
Activity
1
Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2
"Turn" until "Freq" appears in the bottom line.
3
"Press"
Setting:
50H = Mains frequency 50 Hz (factory setting)
60H = Mains frequency 60 Hz