8 Programming
92
8.6 Configuration level 3
At the configuration level 3 parameters of
the Extra Codes and the maths and logics
module are configured.
It is reached from configuration level 2 by si-
multaneously pressing the
h
+
R
keys.
The configuration level 3 covers the follow-
ing parameters:
- relay outputs
- maths and logics module
- interface
- external text
- binary-linked external text
- external stop
- external speed
- event counter
- external scaling
- external report
If a correct coden
umber has been input when calling
up the parameter level, signal acqui-
sition and recording are interrupted
at configuration level 3. The parame-
ters can be viewed and altered.
Signal acquisition, processing and
recording are interrupted.
The relays of the external relay mod-
ule ER8 remain in their current sta-
tus. They are only operated again
when the recorder is at the operating
level or in the basic status.
Display
Edit
Selection/input
with keys
Enter
Continue with
C3-LEVEL
h
forwards
R
backwards