Preparing the microtome for attachment of the ln22, Startup – Leica Biosystems LN22 User Manual

Page 19

Advertising
background image

19

Leica LN22

4.3 Preparing the microtome for

attachment of the LN22

Move the microtome specimen cylinder to the
rear limit stop.

Pivot the magnifier of microscope to the side.

Remove knife holder basis and knife holder.

Remove the specimen clamp.

Unscrew the fixture for specimen clamps.

Lock handwheel (1) in 12-o’clock position (1a).

Instead of the fixture for specimen clamps,
assemble the cryo specimen arm (3) with the
4 shorter screws (standard delivery) (2).

Fig. 11

Rotate the handwheel to place the cryo speci-
men arm (3) at the low end of the cutting
stroke.

Fig. 12

Fig. 10

2

2

4.

Startup

1

1a

3

3