Preparing the microtome for attachment of the ln22, Startup – Leica Biosystems LN22 User Manual
Page 19
Advertising
19
Leica LN22
4.3 Preparing the microtome for
attachment of the LN22
•
Move the microtome specimen cylinder to the
rear limit stop.
•
Pivot the magnifier of microscope to the side.
•
Remove knife holder basis and knife holder.
•
Remove the specimen clamp.
•
Unscrew the fixture for specimen clamps.
•
Lock handwheel (1) in 12-o’clock position (1a).
•
Instead of the fixture for specimen clamps,
assemble the cryo specimen arm (3) with the
4 shorter screws (standard delivery) (2).
Fig. 11
•
Rotate the handwheel to place the cryo speci-
men arm (3) at the low end of the cutting
stroke.
Fig. 12
Fig. 10
2
2
4.
Startup
1
1a
3
3