Lock the handwheel of the microtome

in

12-o’clock position.

Place the specimen holder into the freezing
station, as shown in.

Extra Fine Screw Thread! Do not
tighten too much!

Once the embedding medium is thoroughly fro-
zen, use the special forceps with small radius
(20) to insert the specimen holders into the
cryo specimen arm (2) and tighten with the
clamping screw (3) of the cold bridge (4).